WARNING: When using the Gas Module 3, the maximum sampling rate at the nasal
cannula is 200 ml/min with an Adult/Pediatric water trap and 120
ml/min with a Neonatal water trap. This device should not be used on
patients whose breathing could be impaired by this vacuum flow rate.
WARNING: Connection of the Gas Module 3 exhaust port to the hospital’s waste
gas scavenging system is strongly recommended to prevent exposure
of hospital personnel to the patient’s respiratory sample. Vacuum
(negative pressure) should not exceed 1 mmHg at the Gas Module
Pump Exhaust fitting. Excessive scavenge vacuum may result in
damage to the Gas Module’s internal pump.
WARNING: To ensure data security, connect the equipment to an encrypted
wireless network/network with access control.
WARNING: Monitor Interference may occur if the separation distance from the
disturbance is less than 20 cm.
WARNING: Keep a distance of at least 20cm away from the V 12/V 21 when Wi-Fi
function is in use.