V Series Operating Instructions 18 - 5
Cardiac Output (CO) (Optional) Cardiac Output Dialog
4. Select the Cancel or Accept button.
•The Cancel button closes the dialog without saving the settings.
•The Accept button saves the settings and closes the dialog.
Auto Start When Auto Start is enabled, the monitor
automatically takes the CO
measurement after establishing a
baseline blood temperature. In this
mode, it is not necessary to select the
Start CO button in the CO Run dialog.
When Auto Start is disabled, the clinician
must manually select the Start CO
button in the CO Run dialog in order for
the monitor to begin a CO Run.
On, Off
CO Time Configures the time scale in the CO Run
window.
30 seconds, 60 seconds
CO Scale Configures the temperature scale in the
CO Run window.
0.5 ºC, 1ºC, 2 ºC, 4 ºC
Color Configures the color of the digital data in
the CO digital tile.
16 distinct colors
Factory default color is yellow.
Ready Tone Provides an audible indication that the
monitor is ready to begin a CO run. This
allows the clinician to know a CO run can
be started without actively viewing the
monitor.
On, Off
TABLE 18-1
FUNCTION DESCRIPTION SETTINGS*
* The factory default configuration settings are in bold.

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals