V Series Operating Instructions 29 - 33
Product Specifications Non-Invasive Blood Pressure (NIBP) Performance
Low Alarm Limits:
NOTE: Mean Arterial Pressure (MAP) is defined as:
Mean Pressure 1 = Mean Pressure determined from the oscillometric profile
Mean Pressure 2 = (2 × diastolic + systolic) / 3
Mean Pressure Displayed = (Mean Pressure 1 + Mean Pressure 2) / 2
10. NIBP Static Pressure:
11. NIBP Heart Rate:
12. NIBP Start Pressure Settings and Ranges
The Start Pressure is adjustable throughout the following ranges and set to the following defaults:
Adult: 60 to 200 mmHg in steps of 1 mmHg, default: Off
Pediatric: 40 to 140 mmHg in steps of 1 mmHg, default: Off
Neonate: 40 to 90 mmHg in steps of 1 mmHg, default: Off
Adult: 39 to 150 mmHg in steps of 1 mmHg, default: Off
Pediatric: 39 to 100 mmHg in steps of 1 mmHg, default: Off
Neonate: 28 to 70 mmHg in steps of 1 mmHg, default: Off
Accuracy:
±3 mmHg
Range: 0 to 300 mmHg
Accuracy: Greater of ± 3 BPM or ± 3%
Resolution: 1 BPM
Measurement Range: Adult Mode:35 to 245 BPM
Pediatric Mode:35 to 245 BPM
Neonate Mode:70 to 245 BPM
Invalid Data: Invalid NIBP Heart Rate data displayed as (- -).
TABLE 29-30
PATIENT SETTING
START PRESSURE
RANGE INCREMENT SIZE DEFAULT SETTING
Adult Mode 100-280 mmHg 1 mmHg 180 mmHg
Pediatric Mode 60-180 mmHg 1 mmHg 140 mmHg
Neonatal Mode 40-120 mmHg 1 mmHg 100 mmHg

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals