EasyManuals Logo

Mindray V12 User Manual

Mindray V12
587 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #514 background imageLoading...
Page #514 background image
V Series Operating Instructions 29 - 49
Product Specifications Gases Performance
29.12 Gases Performance
The V 12/V 21 works with Mindray DS Gas Module 3 to measure and monitor inspired and expired
O
2
, CO
2
, N
2
O and five (5) anesthetic agents. For the performance specifications of the Gas Module 3,
refer to Datascope® Gas Module 3 Product Specification.
The ability to properly resolve end-tidal values can be measured by using the set-up described in ISO
80601-2-55:2011 figure 201.101. In short, the method consists of sampling gas from two different
sources connected to an electrically controlled pneumatic valve to permit rapid switching between
the two sources. During the test, the valve is set to switch gas source at a number of frequencies
(simulating the range of specified breath rates) and for each frequency the end-tidal value presented
by the gas analyzer is noted. From a diagram of end-tidal value over frequency, the frequency at
which the gas analyzer is no longer able to resolve end-tidal values according to specification is
identified. This ability to properly resolve end-tidal values is listed in the corresponding AIONâ„¢
Multigas Analyzer technical specification.
1. CO
2
2. O
2
3. N
2
O
Measurement Range: 0–10 vol%; 0-10 kPa; 0–75 mmHg
Measurement Rise Time: 250 ms
Accuracy @ 0–1% of Volume: ±0.1% of Reading*
Accuracy @ 1–5% of Volume: ±0.2% of Reading*
Accuracy @ 5–7% of Volume: ±0.3% of Reading*
Accuracy @ 7–10% of Volume: ±0.5% of Reading*
Gas Cross Effects: <0.3%, If <0.1%, 0.0 is displayed
Threshold: 0.1vol%; 0.0% is displayed if value <0.1%
Respiration Rate Breath Detection: 1.0% change in CO
2
level
Measurement Range: 2 to 100 BPM
Measurement Range: 0–100 vol%
Measurement Rise Time: <400 ms at 200 ml/min
Accuracy @ 0–25% of Volume: ± 1% of Reading*
Accuracy @ 25–80% of Volume: ± 2% of Reading*
Accuracy @ 80–100% of Volume: ± 3% of Reading*
Gas Cross Effects: <0.2% of reading for N
2
O or CO
2
; <1% of reading
for anesthetic agent
Measurement Range: 0–100% N
2
O
Measurement Rise Time: <250 ms
Accuracy @ 0–20% of Volume: ±2% of Reading *
Accuracy @ 20–100% of Volume: ±3% of Reading *
Gas Cross Effects: <0.1% of reading with CO
2
, O
2
or any anesthetic
agent

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals