V Series Operating Instructions 29 - 51
Product Specifications SvO2 Output Performance
29.13 SvO
2
Output Performance
The VDI module has the capability to interface with the Edwards Vigilance®, Vigilance II, the Vigileo™,
and the EV1000â„¢ monitors in Adult and Pediatric Mode. For the performance specifications of the
Edwards monitors, refer to its operator’s manual or contact Edwards Lifesciences Corporation for
assistance.
SvO2 Alarms
The VDI module is equipped with High and Low SvO2 alarms. Alarms can be triggered when
measured rate is equal to or greater/smaller than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
29.14 Analog Output
V 12/V 21 in conjunction with:
• VPS
1. ECG Interface
2. IBP1 and IBP2 Interface
3. Defibrillator Sync Interface
Adult: Off, 20 to 99 in steps of 1, default Off
Pediatric: Off, 20 to 99 in steps of 1, default Off
Adult: Off, 0 to 60 in steps of 1, default Off
Pediatric: Off, 0 to 60 in steps of 1, default Off
Bandwidth (-3 dB referenced to 10 Hz): 0.5 Hz to 100 Hz
Propagation Delay (Delay of QRS complex): Less than 25 msec
Sensitivity: 1 V/mV ±10% (referenced to 10 Hz).
Pacer Displayed with ECG lead: I,II
Pacer Enhancement: Pacer is summed at the output when Pacer enhancement mode is
turned ON.
Pacer amplitude signal is a minimum of 2.5v.
Pacer width is10 ms, with a 5% tolerance. (for single pacer only)
Pacer rise and fall times are100 µs, maximum.
Bandwidth (-3 dB referenced to DC): DC to 15 Hz
Propagation Delay: Less than 25 msec
Sensitivity: 1 V/100 mmHg ±2%
Maximum delay between the QRS peak and the rising edge
of the defibrillator sync pulse:
35 msec
Minimum Output Amplitude: 2V into a 5 Kohm load
Output Pulse Width: 2 msec to 7 msec