Addendum to V Series Operating Instructions 1 - 5
Continuous Cardiac Output (CCO) & Continuous Hemodynamics (Optional) CCO/Hemo Configuration
SV/SV STAT
Unit of Measure: ml/beat
Resolution: 1 ml /beat
Invalid data: “- -
High Limit Off, 20
to 300 Off, 20 to 300 N/A
Low Limit Off, 0 to 140 Of
f, 0 to 140 N/A
SVI/SVI STAT
Unit of Measure: ml/beat/m
2
Resolution: 0.1 ml/beat/m
2
Invalid data: “- -
High Limit Off, 20.0 to
200.0
Of
f, 20.0 to 200.0 N/A
Low Limit Off, 0.0 to 100.0 Of
f, 0.0 to 100.0 N/A
EDV/EDV STAT
Unit of Measure: ml
Resolution: 1 ml
Invalid data: “- -
High Limit Off, 40
to 800 Of
f, 40 to 800 N/A
Low Limit Off, 0 to 400
Of
f, 0 to 400 N/A
EDVI/EDVI STAT
Unit of Measure: ml/m
2
Resolution: 0.1 ml/m
2
Invalid data: “- -
High Limit Off, 20.0 to
400.0
Of
f, 20.0 to 400.0 N/A
Low Limit Off, 0.0 to 200.0 Of
f, 0.0 to 200.0 N/A
RVEF/RVEF STAT
Unit of Measure: %
Resolution: 1%
Invalid data: “- -
High Limit Off, 20
to 100 Of
f, 20 to 100 N/A
Low Limit Off, 0 to 60
Of
f, 0 to 60 N/A
TABLE 1-1
PARAMETER
ALARM
LIMIT ADULT PEDIATRIC NEONATE
* The factory default configuration settings are in bold.

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals