3.4.3 Max. Peak Hold Mode
In this mode, the instrument holds the maximum value in the series of varying measured
values.
Press the [MODE] key twice in the normal mode. "Max" blinks in the display.
Press the [PRESET/SET] key to switch to Max hold mode ("Max" stops blinking).
When the spindle moves, the maximum value is held ("H" appears).
Press the [SET] key to release the hold, display the current position, and start
measuring a new maximum value.
difference of Max,Min,TIR hold mode.
10.00
0.00
5.00
-5.00
a
c
b
d
Path 0
a
b
c
d
Max mode 0.00
5.00
10.00
Min mode 0.00
-5.00
TIR mode 0.00
5.00
10.00
15.00
If max. hold mode is entered from the ABS
system, press the [ZERO/ABS] key for 2 seconds or longer to set the position being
held to zero. The instrument can be used for comparative measurement.
If max. hold mode is entered from the INC system, press the [ZERO/ABS] key for 2
seconds or longer to set the current position to zero.
difference of after Zero set (at Max. hold mode)
operation
[ZERO/ABS]>2sec [PRESET/SET]
ABS 0.00
H
5.00
H
0.00 -3.00
H
2.00
INC 0.00
H
5.00 0.00 0.00
H
5.00
5.00mm
0.00mm
7.00mm
2.00mm
3.4.4 Min. Peak Hold Mode
In this mode, the instrument holds the minimum value in the series of varying measured
values.
Press the [MODE] key three times in the normal mode. "Min" blinks in the display.
Min. hold mode operations are done by the same procedure as max. hold mode.
3.4.5 TIR (Run-Out) Measurement Mode
In this mode, the instrument holds the run-out width in the series of varying measured values.
Only this mode has the same operation in both the ABS and INC systems.
Press the [MODE] key four times in the normal mode. "TIR" blinks in the display.
Press the [PRESET/SET] key to set the display to zero ("TIR" appears).
When the spindle moves, the run-out width is held ("H" appears).
Press the [SET] key to release the held value and start measuring a new run-out
width value.