3.5 AnalogDisplay
An analog scale and pointer in the from of dial gage are continuously displayed at the top of
the instrument’s LCD. In tolerance determination mode, the limit positions blink in this area.
In max./ min. hold mode or TIR mode, the maximum and minimum positions blink in this
area.
3.5.1 Switching the Display Range
The analog display range can be switched to prevent pointers going too high or low.
• Press the [RANGE/→Adj.←] key a short press to toggle the display range.
Resolution Swicthing the display range(loop)
0.001mm
0.02(mm) →0.04 →0.1 →0.2 →0.4
0.01mm
0.2(mm) →0.4 →1 →2 →4
.00005”
0.001(”) →0.002 →0.004 →0.01 →0.02
.0005”
0.01(”) →0.02 →0.04 →0.1 →0.2
.0001”
0.002(”) →0.004 →0.01 →0.02 →0.04
.001”
0.02(”) →0.04 →0.1 →0.2 →0.4
3.5.2 Pointer Centering
When a pointer is out of the analog display's range, use this function to shift the display so that
the pointer is centered. This has the same effect as adjusting the bezel of a dial gage to the
desired scale.
• Press the [RANGE/→Adj.←] key for 2 seconds or longer to center the pointer.
NOTE
The pointer position is adjusted so that the current measured value is in the center.
e.g., In case of judgement tolerance for the 10.000±0.02 (Upper & lower limit:10.020&9.980)
0.000
<
10.004
O
10.004
O
10.004
O
[RANGE]<2sec[RANGE]>2sec
0.02 0.02 0.02 0.02 0.02 0.02
0.04
0.04
3.6 SwitchingtheCountingDirection
By default, the instrument takes the positive direction to be the direction the spindle moves in
when it is pushed in. If desired, this direction can be set as the negative direction.
• To reverse the counting direction, press the [+/-] key in the normal mode ("REV"
appears).
3.7 FunctionLock
•
Pressing the [+/-] key for 2 seconds or longer deactivates all key input except
ON/OFF and hold release (Lock sign appears).
• To reactivate the keys, press the [+/-] key again for 2 seconds or longer.
• To reactivate the keys, press the [+/-] key again for 2 seconds or longer.