5 ERROR MESSAGES & CORRECTIVE MEASURES
Display Description
An ABS data composition error occurred while the spindle was moving at high speed.
Keep on using the instrument since this error does not affect measured values.
*If this error occurs while the spindle is stopped, it could be due to sensor failure. In
this case, contact a Mitutoyo sales office or agency.
The upper limit has been set to a smaller value than the lower limit. Set the limits again
so that the lower limit is less than the upper limit.
Upper limit overflow. Set the upper limit again.
Lower limit overflow. Set the lower limit again.
The display value has exceeded the maximum number of digits which can be displayed.
Set the origin again to a suitable position.
Below are errors which can occur when the Presetter is used. For more information, see the
manual that comes with the Presetter.
Using the Presetter, the lower limit has been set to a higher value than the upper limit.
Alternatively, the preset value or upper limit value were input while the instrument was
waiting for input of the lower limit value, because the lower limit was set first.
Press the [ZERO/ABS] key to release the error, then redo the setting.
Communication error. The number of decimal places in a preset value sent from the
Presetter differs from the number in the instrument's LCD display. Press the
[ZERO/ABS] key to release the error, then redo the setting.
Presetter connection error. Press the [ZERO/ABS] key to release the error, then check
the connections.