4.3TimingChart
0 sT1<500µs
15µsT235µs
270µsT3410µs
230µsT4400µs
IMPOTANT
Use only the output cable specified by Mitutoyo. Use of incompatible or
deteriorated cables may result in data output failure.
Before outputting data, read the manual that comes with the data processor
carefully to ensure correct operation.
Data output may be disabled if an output request (REQ) is received when the
spindle is in motion, or if the output request interval is too short.
CAUTION
*1: To make an output request (REQ), hold the REQ signal at "Low" until "CK"
is output. Return it to "High" before the final (52nd) "CK" bit is output.
REQ
DATA1
CK
T1 T2 T3 T4
*1
1bit
2bit
3bit

52bit
4.4 UsingtheDigimaticPresetter
The instrument can use the optional Digimatic Presetter (No. 543-003; sold separately) to set
preset values and limit values externally. The operation method is described in the manual for
the Presetter. There are some differences in the instruments operation when the Presetter is
used, described below. Read these points when using the Presetter.
4.4.1 Setting Limit Values from the Presetter
Once the Presetter has set the upper limit in the instrument, the
instrument waits for input of the lower limit. Set the lower limit using
the Presetter.
<
When the lower limit has been set, the instrument enters tolerance
determination mode.
When the instrument is waiting for input of the lower limit, pressing
the Presetter's [PASS] key cancels the limit settings and returns to
the normal mode.
MANUAL No.99MAH001B2