Sterilizers SUPREME - FUTURA - CLASSIC (17-22-28)
- TECHNICAL MANUAL
EN
SERVICE MENU
55
SQ11 CHECK EV1/EV3 FOR PROPER CLOSING
- close the door
- from SOLENOID VALVES menu:
- close EV1
- open EV4
- go to the OTHER
COMPONENTS menu:
- note the reading P” of
chamber pressure
- activate VP vacuum pump
If EV1/EV3 are closing properly,
chamber pressure value P
will not change.
With the door open, proper closing
of EV1/EV3 is confirmed when
there is no suction from:
- steam outlet, top/rear section of
chamber for EV1
- chamber filter coupling for EV3
- de-activate the components after
completing the test
SQ12 CHECK FOR EV1 OPENING
- close the door
- from SOLENOID VALVES menu:
- close EV1 and EV4
- open EV3
- go to the OTHER
COMPONENTS menu:
- note the reading P” of
chamber pressure
- activate VP vacuum pump
If EV1 is open, chamber pressure
value P” will decrease.
With the door open, EV1 opening
can be confirmed by checking the
effect of suction from the steam
outlet, top/rear section of chamber.
- de-activate the components after
completing the test