Preparing the Software
Entering software preferences
Selecting instrument settings
Defining templates, reduction parameters, and display
parameters
Confirming hardware and software setup
Running the Experiment
Initiating the operation (detection or fluidics plus detection)
Saving the data file
Analyzing the Data
Modifying the template or parameters as desired
Saving the data file
Analyzing the data
Exporting data to another software application as desired
Theory of Operation
This section includes the following topics:
Instrument Design, see page 33
Assay Read Types, see page 38
Instrument Design
Fluidics
The instrument is designed with a fluidics module that transfers liquids to
the assay plate during a fast kinetic read or before an experiment.
The fluidics module incorporates an 8-channel or 16-channel pipettor
that automatically changes tips and transfers reagents to the plate that is
read in the system. Pipette height and dispensing rate are adjustable. The
instrument can add reagents within milliseconds of a column being read,
enabling fast kinetic assays of transient responses.
As many as three compounds can be transferred from columns in a
compound plate to a single column in an assay plate, at different points
during or before the total read time.
FlexStation_manual.book Page 33 Tuesday, June 25, 2013 10:22 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Molecular Devices FlexStation 3 and is the answer not in the manual?

Molecular Devices FlexStation 3 Specifications

General IconGeneral
BrandMolecular Devices
ModelFlexStation 3
CategoryCard Reader
LanguageEnglish