0112-0127 B 155
B
Performance Specifications
Thermal specifications for microplates used in the FlexStation 3
instrument apply to flat-bottom microplates with isolated wells. All other
microplate specifications apply to standard 96-well polystyrene
flat-bottom microplates.
When pathlength compensation is applied to microplate absorbance
measurements, the solution volume in the microplate well is between
100 μL and 300 μL.
Technical specifications are subject to change without notice.
Table B-1: Fluidics Specifications for the 8-Channel Pipettor
Item Specification
Volume range 1 μL to 200 μL
Dispensing precision 8% CV at 5 μL
2% CV at 50 μL
Maximum dispense rate 208 μL per second
Table B-2: Fluidics Specifications for the 16-Channel Pipettor
Item Specification
Volume range 1 μL to 30 μL
Dispensing precision 3% CV at 10 μL
5% CV at 1 μL
Maximum dispense rate 52 μL per second
FlexStation_manual.book Page 155 Tuesday, June 25, 2013 10:22 AM