6
[3] Installation
WARNING
The equipment must be installed by a qualified and trained professional
specified by J.Morita Corp. according to the conditions specified in the
Installation Standards Manual.
>>Otherwise, there is the dan
er of an accident or equipment dama
e.
The unit must be fastened and fixed to the floor securely as specified in
the installation manual.
>>Otherwise, the chair could fall over resultin
in in
ur
or dama
e.
The equipment MUST NEVER be connected to an electrical supply source
which is not the specified voltage, current, frequency and power.
Each unit must have its own devoted electrical circuit connection and
should not share a circuit with other units or other types of equipment.
>>Otherwise, there is the danger of shorts or over heating resulting in injury or
dama
e.
Never splash the equipment with water or place it in an area where it could
get wet.
>>This would create the danger of shorts or overheating resulting in injury or
dama
e.
Do not install the unit in an area where chemicals are stored or where gas
could be released.
>>This would create the danger of fire or electrical shock resulting in injury or
dama
e.
Do not route the power supply cord, the earth lead or the foot control cord
through areas such as doorways and halls where they could be damaged.
>> This would create the danger of shorts and electrical shocks resulting in
in
ur
or dama
e.
CAUTION
Do not tilt the equipment or place it in an unstable position; do not expose
it to excessive vibration of physical bumps or shocks.
>> This could cause it to malfunction or dama
e it.
Do not install the equipment in areas where it will be exposed to excessive
atmospheric pressure, humidity, temperatures, wind, direct sunlight, dust,
salts or sulfur compounds.
>> Any of the above conditions could cause the equipment to malfunction or
dama
e it.
SIGNO G10 must be installed in a room appropriate for dental treatment.
1)Temperature: 10~40
℃
2)Humidity:30~75%RH(without condensation)
3)Pressure:700~1060hPa
>> Otherwise it ma
fail to perform optimall
, and it could malfunction as well.