8. Clean the Drain Trap
CAUTION
Clean the drain trap at least once a month.
>>Otherwise, it could cause performance deterioration, breakdown, or clogging
of pipes.
After cleaning, make sure the drain trap filter and cap are properly and
securely installed.
>>Otherwise, drain water could leak resultin
g
in dama
g
e or a malfunction.
Clean the drain trap at least once a month or whenever the basin does not seem to drain
properly.
Follow the procedure below to do this.
(1) Raise the chair about 100mm and above its lowest position
and turn off the main switch.
Place a bucket in the lower part of the unit to
catch drain water filled in the cap.
(2) Open the door on the unit and unscrew the cap of the
trapping part.
The drain trap filter is built in the cap.
Give it a turn to the left slightly, then a lock is released. Take
off the cap straight down so that it is not inclined.
ADVICE
Cap
<Inside of the unit>
OFF
Main Switch
(Lamp off when OFF)