Test specifications for ENCLOSURE PORT IMMUNITY to
RF wireless communications equipment
Test
Frequency
(MHz)
Band
a)
(MHz)
Service
a)
Modulation
b)
Maximum
Power
(W)
Distance
(M)
IMMUNITY
TEST LEVEL
(V/m)
385
380 - 390
TETRA 400
Pulse
modulation
b)
18Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
C)
± 5kHz deviation
1 kHz sine
2
0.3
28
710
704 - 787
LTE Band 13,
17
Pulse
modulation
b)
217 Hz
0.2
0.3
9
745
780
810
800 - 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b)
18 Hz
2
0.3
28
870
930
1720
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4,25; UMTS
Pulse
modulation
b)
217 Hz
2
0.3
28
1845
1970
2450
2400 - 2570
Bluetooth,
WLAN,
802. 11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b)
217 Hz
2
0.3
28
5240
5100 - 5800
WLAN 802.11
a/n
Pulse
modulation
b)
217 Hz
0.2
0.3
9
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and
the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by
IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
Basic performance: none
Intended behavior
Cable length
Three-way Syringe
Output:
Water and Spray: 30mL / 15sec., minimum
Main tube: about 1.5 m
Foot control: about 700 mm
Vacuum Syringe Performance: Volume: 90 L / min., minimum
It depends on the vacuum motor.
(Reference value)
Volume: 140 L / min., minimum
(with EV-12 Type)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Morita SIGNO G10 II TU101 and is the answer not in the manual?

Morita SIGNO G10 II TU101 Specifications

General IconGeneral
BrandMorita
ModelSIGNO G10 II TU101
CategoryDental equipment
LanguageEnglish

Related product manuals