USER MANUAL NOVAcompact
MRU GmbH, D-74172 Neckarsulm
17 / 50
Settings measurement
Temperature unit
°C, °F
Change the unit for temperature in all
screens
Pressure unit
Pa, hPa/Pa, hPa,
kPa/Pa, kPa, mbar,
mmH2O, cmH2O,
inH2O, mmHg, inHg,
PSI,
Change the unit for pressure in all
screens. The meaning of hPa/Pa and
kPa/Pa is that the instrument performs a
dynamic change of unit depending on
the absolute value of pressure.
Draft unit
Pa, hPa/Pa, hPa,
kPa/Pa, kPa, mbar,
mmH2O, cmH2O,
inH2O, mmHg, inHg,
PSI,
Change the unit for pressure in all
screens. The meaning of hPa/Pa and
kPa/Pa is that the instrument performs a
dynamic change of unit depending on
the absolute value of pressure.
Core flow search
ON / OFF
Core flow search before start of each
flue gas measurement: activated or
deactivated
Combustion calculation
ON / OFF
If the combustion calculation is switched
off following items will be changed:
- no fuel types, respectively always
"Sample Gas"
- no measuring values losses, Verluste,
ETA, ETAcond, Dev. point
- no measuring value CO2, except it will
be measured
- no measuring values CO/NO/,,,. in
[mg/kWh]/[mg/MJ]
- no fuel type will be shown on the menu
bar and print-out
Input soot & T-Boiler
ON / OFF
Input soot and T-boiler values for print-
outs and/or storage: activated or
deactivated
Annular-gap test
ON / OFF
Annular-gab measurement: activated or
deactivated
Explanation for “Helping hints”:
Some helpful hints which are very useful for an inexperienced user but are not needed by
experienced users, can be activated or deactivated. The following hints will be affected:
“Zeroing finished, Sensors are ready. Analyzer ready for measurement.”
“Reminder! Charge batteries at regular intervals!”
“Measurement stopped/started.”