10. Technical specifications
Measured values
NOVAcompact
O
2
measuring range
0 - 21,0 Vol-%
Accuracy
± 0,2 Vol-% abs.
Reaction time T90*
< 20 s
CO (H
2
komp.) measuring range
0 - 4.000 ppm, overload up to 10.000 ppm
Accuracy
± 10 ppm or** 5 % of the measured value up to 4.000 ppm or**
10 % of the measured value up to 10.000 ppm
Reaction time T90*
< 40 s
CO high measuring range
0 - 4.000 ppm, overload up to 20.000 ppm
Accuracy
± 100 ppm or** 5 % of the measured value up to 4.000 ppm or**
10 % of the measured value up to 20.000 ppm
Reaction time T90*
< 40 s
NO measuring range
0 - 1.000 ppm, overload up to 5.000 ppm
Accuracy
± 5 ppm or** 5 % of the measured value up to 1.000 ppm or**
10 % of the measured value up to 5.000 ppm
Reaction time T90*
30 s
Flue gas temperature T
A
Measuring range
0 - 650 °C with high grade steel probe pipe
Measuring range
0 - 1.100 °C with Inconel probe pipe
Accuracy
± 2 °C 200 °C
1 % of the measured value > 200 °C
Ambient air temperature T
L
Measuring range
0 - 100 °C
Accuracy
± 1 °C
Draft
Measuring range
± 100 hPa
Accuracy
± 0,02 hPa or 1% of the measured value
Differential pressure
Measuring range
± 100 hPa
Accuracy
± 0,02 hPa or 1% of the measured value
Max suction range gas pump
150 hPa
Typical gas flow
60 l/h