Attention! This function is not suitable for back up or for the transfer to another analyzer
because the exported file can not be imported again!
The created file has the name ‚EMIxxxxx.csv’, in which the xxxxx are continuing 5 digit numbers
with leading zeros.
The created file has a column header with the following information: Site number, Date/Time,
Measuring program name, Fuel type, CO2max, O2reference, and all measured values that the
analyzer can measure as well as the soot numbers, Derivate and T-Boiler.
Example:
Export of Differential Pressure Measurements
The same function as Export of Flue Gas Measurements only the file name is different.
The created file has the file name “DDMxxxxx.csv“, in which the xxxxx are continuing 5 digit
numbers with leading zeros.
The created file has a column header with the following information: Site number, Date/Time, as
well as 4 saved pressure measurements.
8.5. Measurements in the Data Memory
View measurements
In the menu item „View measurements“you can inspect the stored measurements. After
selection of this item you receive first an overview of the number of the stored measurements
according to measuring type.
Select flue gas measurement or another measuring type.
Then you receive first a page with context information to the stored measurement. Scroll
with the arrow keys by the context information of the stored measurements.
With F2 = "measured value" are displayed the measured data of the stored measurement in
detail, available in 3 measuring value pages, as they are defined in the measuring value
window.
With ESC you return to the context information of the measurement.
You have the possibility to display only those data that are assigned to a single site:
either F1 = „this site“ , while a measurement of the desired site is displayed.
With F1 = „all sites“ you cancel this filter again.
or while you select with the menu key the function "search a site" and execute, as
described in the chapter site administration.
Delete measurements
You are able to
delete single measurements, while they are displayed press the key
F3 = "delete" .
or delete all measurements of a measuring type.