After the Bump Test
After all installed sensors pass the Bump Test, the symbol displays on the
MEASURE page. This symbol appears on the:
- colour display in the upper feature bar
- monochrome display in the lower-right corner.
If any sensor was not bump tested, or fails the Bump Test, the symbol does not
display.
The colour display:
- temporarily shows the symbol at each gas reading for successfully bump test-
ed sensors
- symbol is then replaced by the present gas reading.
The monochrome display does not show symbols for individual gas readings.
The symbol shows for 24 hours after the Bump Test.
If a sensor fails the Bump Test, calibrate the device as described in chapter 3.8.
3.8 Calibration
The ALTAIR 5X can be calibrated either manually using this procedure or automat-
ically using the GALAXY test stand. Refer to chapter 7.5.
The use of the demand regulators listed in chapter 6 is recommended.