Operation
ALTAIR 5X/ALTAIR 5X IR
38
GB
3.4 Device Setup
The device allows the user to access and modify the following parameters through
direct button interface:
- Calibration Options
- Alarm Options
- Instrument Options
These menus can be accessed only from the measure page by pressing and hold-
ing the and buttons simultaneously until you are prompted for a password.
The operation is as follows:
(1) Turn the device ON and wait until the measure page appears.
(2) Simultaneously press and hold the and buttons for approximately five
seconds.
The default password is "672".
(3) Enter the first digit by pressing the or button and confirm with the ON/
OFF button.
The cursor jumps to the second digit.
(4) Enter the second as well as the third digit.
Incorrect password: device returns to the Main Page.
Correct password: user can set the device options.
The password can be changed with a PC through the MSA Link software.
The following Options are available by pressing the and buttons:
- Calibration Options - see chapter 3.4
- Alarm Options - see chapter 3.4
- Instrument Options - see chapter 3.4
Monochrome display Colour display
PASSWORD PASSWORD
000 000