11. Reconnect the main operating power, and reconnect the
WHITE wire to the Toxgard II Monitor internal backup
battery (if so equipped).
"
WARNING
Verify the proper operation, sound and loudness of the Banshee
Horn; otherwise, the horn cannot properly alert the user.
Autocalibration Option
The Toxgard II Monitor is available with an autocalibration option
This user-selectable option allows periodic, unmanned gas
calibration of the system.
This option is not available for units with sampling options, nor
is it available for monitoring of the following target gases:
chlorine
chlorine dioxide
hydrogen chloride
nitrogen dioxide.
For installation details, refer to the autocalibration Instruction
Manual (P/N 815379).
Duct-Mounted Remote Sensor Installation
Refer to Appendix A for this option.
Chapter 1, Set-up
1-32