calibration, it will be held at the lockout value
for an additional two minutes.
c. If a CAL FAULT flag appears on the unit (FIGURE 3-11),
this indicates:
• An unsuccessful attempt to zero or calibrate the
Toxgard II Gas Monitor
• The Toxgard II Gas Monitor is operating with the
calibration parameters defined before the calibration
was attempted.
To extinguish the CAL FAULT flag, a complete calibration
procedure must be performed.
• See Troubleshooting Guidelines found in Chapter 5.
The Toxgard II Gas Monitor allows automatic zero adjustment
only within a pre-defined range. It cannot make any
corrections outside this range, such as when an empty or
wrong cylinder of gas is applied or failure to begin gas
flow within the allotted 30-second countdown occurs.
Figure 3-10.
Calibration End Display
Figure 3-11.
Calibration Fault Flag
Chapter 3, Calibration
3-11