3. Point the Calibrator or Controller at the Toxgard II Gas
Monitor display (FIGURE 1-2); press the ZERO button or
push the right-hand internal CAL push-button (FIGURE
1-1) until the word "ZERO" appears; then, release.
NOTE: The zero or calibration process can be aborted at
any time during the 30-second countdown interval;
simply press any button on the Calibrator or
Controller while aiming it at the unit or by pushing
the internal push-button.
The 30-second countdown interval is omitted
for oxygen units. It is electronically zeroed.
The display shows:
A countdown from 30 to 0 seconds
APPLY ZERO GAS (FIGURE 3-9)
4. After the 30 second countdown:
The display alternates between "CAL" and a value (for
example: 0 PPM). This value is the actual reading of the
gas concentration the sensor is detecting.
Once the gas value on the display is stable for 15
seconds, the alternating display stops. If the calibration is
successful, the display will show END and 0
(FIGURE 3-10).
a. If using the zero cap: remove it.
b. If using a zero gas cylinder:
1) Turn OFF the gas flow by turning the flow
controller knob.
2) Remove the tubing from the SensorGard.
If the calibration signal is enabled during
Figure 3-9.
Apply Zero Gas Flag
Chapter 3, Calibration
3-10