Configuring the job computer
Configuring the nozzles - for field sprayers with pressure sensor regulation
1. - Press for the icon to appear in the “User data” area.
2. Enter the intended application rate in the field.
3. Enter the pressure range for the nozzle colour.
4. In the “Results” area, you can see the speed at which the application rate can be achieved.
Calibrate the nozzles
“Calibration” screen
Nozzle output at 3 bar - Result of the
calibration
Current nozzle output. When you change the
pressure, the newly calculated nozzle output
appears.
For non-standard nozzles, the calibration must
be repeated at a different pressure. This
pressure should be about as high as the
intended working pressure.
Spray liquid or spray liquid residues
Danger of poisoning or chemical burning
◦ Clean the spray liquid tank thoroughly prior to calibration. The field sprayer must be free of
spray liquids or spray liquid residues.
◦ Use only clear water during calibration.
◦ Wear required protective equipment.
Calibrate a standard nozzle as follows:
The tank is filled with clear water.
There are no spray liquid residues in the tank or the pipelines.