EasyOne Pro™ / EasyOne Pro™ LAB 101
Hygiene, Cleaning, Maintenance, Disposal
Maintenance
Before each use
Before each use, visually inspect the device, the cables, the tubing and the
sensor for signs of mechanical damage. Particularly inspect the gas cylin-
der(s) and fittings.
If you detect damage or impaired functions that may adversely affect the
safety of the patient or user, do not use EasyOne Pro™ before it has been
repaired.
Once a year
The following parts must be replaced once a year (see "Start-up and Initial
Preparation" on page 20):
the filter pack
the gas supply tube
the one-way valve
the overpressure valve.
All parts are included in the "Yearly Maintenance Kit" 3000-50.50SP (see
"Order Information" on page 134).
The filter pack at the back of the device consists of the air inlet with filter
and a Nafion tube to dry the sampled gas. Open the screw on the right to
replace the filter pack (Fig. 7-1).
Fig. 7-1 Screw to replace the filter pack