Performing a Test
52 EasyOne Pro™ / EasyOne Pro™ LAB
Edit the values, if necessary, and click Confirm >> to close the window.
Warning
Erroneous Measurement Results—Incorrect entries may lead
to erroneous test readings and incorrect system
interpretations. The specified measuring accuracy is ensured
only when all entries are correct.
The inspiratory and expiratory FVL test acquisition screen appears.
Measurement
Prepare the patient (see "Preparations" on page 46) and explain the test
maneuvers to him or her:
take the spirette™ into the mouth with the lips sealing around the
spirette™, taking care not to block its opening with the tongue or bite
down excessively on the spirette™
breathe calmly
fill lungs completely
exhale as hard and fast as possible
continue blowing out until the lungs are completely empty
inhale as hard and fast as possible.
Fig. 3-14 Window showing the ambient conditions
a Conversion factors
b Box for entry of the humidity
c Box for entry of the ambient temperature
d Box for entry of the altitude (atmospheric
pressure)
e Click to recalculate the BTPS factor (if the
ambient conditions were modified)
f Click to confirm and accept the values