EasyManuals Logo

NDD easyone pro User Manual

NDD easyone pro
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
EasyOne Pro™ / EasyOne Pro™ LAB 53
Performing a Test
Conduct the tests as described in section "Measurement" on page 48.
Note
The volume-time curve (Fig. 3-16) does not display the
forced inspiration at the end of the maneuver.
Bronchodilation (Post-Tests)
The Post test is usually performed to determine the patient's response to
bronchodilating asthma medication. This is done by administering a bron-
chodilator to the patient after he or she has performed a spirometry test
(Pre-Test). Approximately 10 to 20 minutes after the medication (when
bronchodilator shows effect, refer to medication labelling) a second spirom-
etry test is performed. After the test the Pre and Post results are compared.
Post tests can only be added to previous tests performed on the same day.
Select the patient.
Select the test, then click Add Post to close the window.
The window with the patient's existing tests appears. With
(Fig. 3-15
left) you display the detailed results of the Pre tests (Fig. 3-15 right).
Conduct the tests as described in section "Measurement" on page 48.
As soon as three acceptable trials have been performed, the message Ses-
sion complete! appears and a quality grade from A (optimal) to F is dis-
played (see "Quality Messages and Quality Grades" on page 84).
Fig. 3-15 Detailed test results