5.3 Explanation of Commands
DF1906
・[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <amplitude> VPP|VRMS
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Explanation: Sets/queries the output voltage.
Parameter: Ou tput voltage [Vp-p] or [Vrms]
0.000 (0 Vp-p) to 10.000 (10 Vp-p)
0.000 (0 Vrms) to 3.535 (3.535 Vrms)
Response: Returns the setting voltage in [Vp-p].
Nu meric data is transferred as ASCII byte in NR2 format.
ex) VO LT ?
10.000
Remark: The setting of voltage in Vrms is only valid when the sine wave is selected for the
waveform type, but when a waveform other than the sine wave is selected, an
error occurs.
・
[:SOURce]:VOLTage:UNIT {VPP|VRMS}
[:SOURce]:VOLTage:UNIT?
Explanation: Sets/queries the display unit of the output voltage.
Parameter: Ou tput voltage display unit
VPP
VRMS
Response: Returns VPP or VRMS.
ex) VO LT:U NI T ?
VPP
Remark: The Vrms setting is only valid when the sine wave is selected for the waveform
type, but when a waveform other than the sine wave is selected, an error occurs.
・[:SOURce]:VOLTage[:LEVel][:IMMediate]:OFFSet <offset>
[:SOURce]:VOLTage[:LEVel][:IMMediate]:OFFSet?
Explanation: Sets/queries the DC offset voltage.
Parameter: DC offset voltage [V]
-5.000 (-5 V) to +5.000 (+5 V)
Response: Returns the set DC offset voltage in [V].
Nu meric data is transferred as ASCII byte in NR2 format.
ex) VOLT:OFFS?
5.000