ValveMate 7197PCP-2K Controller
Variable Range Description
Info n/a Select to view information about the current screen, including the
range limits for settings.
RPM 10–150 (RPM) Sets the motor speed in RPM; for guidance on setting the RPM, refer
to “Maximum Motor Speed Based on Viscosity” on page47.
Reverse % 0–200 (%)
(adjustable in
increments of 1%)
Based on the percentage of rotation, sets the suck-back to reverse the
motor at the end of a dispense cycle to prevent drooling.
Correction Factor 0.1–2.00
(adjustable in
increments of 0.01)
Because rotors and stators may not be perfectly matched, the
Correction Factor linearly scales the output to ensure that the
expected amount is deposited every time.
Analog On /
Analog Off
n/a Select Analog On to use the “RPM: 10V” and “RPM: 0V” fields to
change the motor speed on-the-fly. When Analog Off is selected, the
“RPM:10V” and “RPM: 0V” fields are disabled.
Analog RPM: 10V 10–150 Scales the output RPM linearly from 0–10V based on the input
analog voltage (pins 12 and 13 of the I/O port; refer to “I/O Port Pin
Assignments and Wiring Diagrams” on page44 as needed)
Analog RPM: 0V 10–150
Programming (continued)
Line Programs
Use a Line program to dispense a continuous line of material. When a Line program is run, the pump dispenses for
as long as the dispense cycle is initiated. You can enable Analog On to fine-tune the motor speed while running a
Line program.
1. Navigate to the Line program screen.
2. (Touchscreen interface only) Select the pump button to toggle between the Pump 1 and Pump 2 screens.
3. Select the ENABLE LINE PROGRAM radio button.
NOTE: If you don’t enable the program, the system will not save any entered settings.
4. Enter the desired settings, referring to the table below for detailed information on each variable.
5. Select SUBMIT to save the settings. The variable table shows the saved settings.
6. To save the values you entered as a program in the Program Library, refer to “Saving a Program to the Program
Library (Save Screen)” on page34.
Line program screen, touchscreen interface
Line program screen, web interface