EasyManua.ls Logo

OHAUS TP4KD - Page 104

OHAUS TP4KD
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
51
CommandCommand
CommandCommand
Command
CharacterCharacter
CharacterCharacter
Character
DescriptionDescription
DescriptionDescription
Description
PP
PP
P Print display data
When “numeric only” display
data is selected for output in
the RS232 menu, the Mode
field is not output.
xSxS
xSxS
xS Set stable data only printing (set x = 0 for OFF, or x = 1 for ON).
TT
TT
T Same effect as pressing tare button
VV
VV
V Print EPROM version
xZxZ
xZxZ
xZ Set Auto Zero to “x”,
where x = 0 to 3 (see table).
x%x%
x%x%
x% Downloads reference weight “x” for percent mode. “x” must be in grams.
Command is ignored if percent mode is disabled. If percent mode is enabled,
balance will automatically switch to percent mode display.
x#x#
x#x#
x# Downloads average piece weight “x” for parts counting mode. “x” must be in
grams. Command is ignored if parts counting mode is disabled. If parts
counting is enabled, balance will automatically switch to parts count display.
Esc LEsc L
Esc LEsc L
Esc L Prints listing of Setup and Print menu settings.
Esc REsc R
Esc REsc R
Esc R Resets Setup and Print menus to factory defaults.
CAUTION: This will reset RS232 configuration.
Esc SEsc S
Esc SEsc S
Esc S Save current settings.
Field: Weight Mode Stab CR LF
Length: 9 1 5 1 1 1
Same as ?
command
Displayed weight sent right justified
w/lead zero blanking.
Nine characters include:
decimal point (1)
weight (7 max))
polarity (1): blank if positive
“ - ” if negative
Field: Model # EPROM # CR LF
Length: 7 15 1 1
0 = OFF
1 = .5 d
2=1 d
3=3 d
Balance Model
"98101-XX Sr*XX.X"