53
Error CodesError Codes
Error CodesError Codes
Error Codes
The following list describes the various error codes and which can appear on the
display and the suggested remedy.
Data ErrorsData Errors
Data ErrorsData Errors
Data Errors
0.0 Transient error (hardware error, probably static discharge). If error persists,
the balance must be serviced.
Tare ErrorsTare Errors
Tare ErrorsTare Errors
Tare Errors
2.0 Balance is unable to stabilize within time limit after taring. Environment is too
hostile or balance needs recalibration.
2.1 Power on load out of specification.
Calibration ErrorsCalibration Errors
Calibration ErrorsCalibration Errors
Calibration Errors
3.0 Incorrect or no calibration weight used for calibration. Recalibrate with correct
weights.
RS232 ErrorsRS232 Errors
RS232 ErrorsRS232 Errors
RS232 Errors
4.0 Bad RS232 frame. Check RS232 menu parameters and correct.
4.4 RS232 buffer is full (if installed). May occur if no printer or computer is
connected to the interface. To clear buffer, turn balance off or enter Print menu
and select END.
4.5 Function is disabled by the Lock switch.
User ErrorsUser Errors
User ErrorsUser Errors
User Errors
7.0 User entry out of bounds.
7.1 Bad percent (%) mode, sample too low.
7.2 Number outside of display capacity.
Over-Under Load ErrorsOver-Under Load Errors
Over-Under Load ErrorsOver-Under Load Errors
Over-Under Load Errors
8.0 Hardware error causing an internal weight signal which is too low. Check if pan
or pan support is off. If not, the balance must be serviced.
8.1 Hardware error caused by an internal weight signal which is too high. Check
load on the platform which may be excessive. If error persists, the balance must
be serviced.