EasyManua.ls Logo

Olympus BH2 Series - B2-BIC Unit Disassembly

Olympus BH2 Series
103 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10-1-4
Disassemble
'B2-BI
Unit
from B2-BIC Unit
by
removing
four
Screws CUK3x6SA.
(See
Fig.
10-4)
Fig.
10-4
10-2
Disassemble
B2-BIC
Unit
as
follows:
10-2-1
Disassemble
Prism
Mount
AB032900
from
Body
AB033000
by
removing
three
Screws
CUK3x8SA.
(See
Fig.
10-5)
Fig.
10-5
-
---
-
-
-
-
-
-
-
- -
-
-.
---
--
-
-,
10-2-2
Remove
P2
LP093900
from
Prism
Mount
CP093900
AB032900:
1
a)
After
loosening
two
Screws
HU3x4SA,
remove
the
Adhesive
OT1028
(Araldite).
(See
Fig.
10-6)
Fig.
106