7.
DISASSEMBLY
PROCEDURE
FOR
BH2-CH
Fig.
7-1
Detach
Knob
A8059800
by
removing
Screw
ACUT.Gx4SA.
(See
Fig.
7-2.)
I-?
Disassemble
the
Stage
Mount
in
the
following
Fig.
7-2
manner:
7-1
-
1
Remove
Stopper
H
K2.6x4SA.
Stopper
H
K2.6x4SA
can
he
uncovered
by
lowering
the
Condenser
Mo'lder.
(See
Fig.
7-1
.)
7-1-2
Remove
the
Condenser
Holder
by
loweriny
it.
7-1
-4
Remove
Nut
AA007700
by
using
Tool
KKAA-
I
I
I
0077,
(See Fig.
7-3.)
7-1
-5
Remove
Tension
Screw
AB026900.
(See
Fig.
7-3.)
7-1-6
Remove
Ball
B3J32,
Pinion
AB026800
and
Washer
AA272600.
/
Fig.
7-3