D
-
3.
DISASSEMBLY PROCEDURE FOR FINE
&
COARSE ADJUST3MQ&T OF
BHS
3-1 Remove the Electrical Base plate from the
Microscope Stand in the procedure given below:
Fig.
3-1
3-1-1 Place the Microscope Stand upside down and
remove three Screws
AB6x25SA.
(See Fig.
3-2.)
3-1-2 While holding the Electrical Base plate
by
hand to prevent
it
from dropping down,
reverse the Microscope Stand.
CAUTIONS.
o
The Electrical Base plate
is
very heavy
and must be handled carefully.
o
While the Microscope Stand is placed up-
side down,
it
is
dangerous to remove the
Electrical Base plate which has protru-
sions comprising internal wiring, etc.
Fig.
3-2
Be sure to remove the base as autlined
abova.
3-19
Whllr
grr~plng tha
Arm
by
hand, remove the
Mlcro8cop~
Bund
upward.
I
Fig.
3-3
-21
-