EasyManua.ls Logo

Olympus BH2 Series - Disassemble Stopper

Olympus BH2 Series
103 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
c)
Remove
Nut
AA782800
by
using
Tools
KC-2010
and
KKAA7828.
{See
Fig.
3-
f
6.)
NOTE:
The
Nut
is
fixed
with
Adhesive.
d)
Extract
Pinion
Mount
AB023600
in
the
direction
indicated
by
arrow
in
Fig.
3-16.
Balls
B1/16
are
removed
together.
el
R
ernove
Screw
AA428500.
(See
Fig.
3-16.)
NOTE:
Tbe
Screw
is
fixed
with
Adhesive.
~~782800'
K
~~~7828
Fig.
3-76
3-3-1
2
Disassemble
right
Coarse
Adjustment
Knob
3PUK
2.6
X
5SA
in
the
following
sequence:
a)
By
removing
three
Screws
3PU
K2.6x5SA,
detach
Knob AB059400.
b)
By
removing Screws
3PUK2.6x5SA
and
two
Screws
S
K2.6x5SA,
detach
Plate
AA783000.
c)
Remove
gear
assembly
23808800,
ZA-
078200
and
2
A078300.
(See
Fig,
3-1
7.)
I
!
Fig.
3-17
3.4
Disassemble
Stopper
ZJ808500
as
follows.
3m4.1
Remove
Stopper
Screw
AAOOl
500.
3-4-2
Remove
Knob
AA785200
by
turning
it
in
the
direction
indicated
by
arrow
in
Fig.
3-18,
fr
$rn
-'-/
\;
!
3-43
Remove
R
ing
AA784100.
(See
Fig.
3-1
8.)
---\
AA784100
Fig.
3-18