Impedance Result Formats
Magnitude: Displays the magnitude of the measured
impedance in Ohms.
Magnitude (dB): Displays the magnitude of the
measured impedance in dBΩ.
Phase (°): Displays the phase of the measured
impedance in degrees.
Phase (rad): Displays the phase of the measured
impedance in radians.
Tg: Displays the group delay of the measured
impedance in seconds (see Gain result equations for
details).
Polar: Displays the measured impedance in a polar
chart.
Real: Displays the resistance of the measured
impedance in Ohms.
Imaginary: Displays the reactance of the measured
impedance in Ohms.
Rs: Displays the equivalent series resistance (ESR) of
the measured impedance in Ω. This value equals the
Real part of impedance.
Ls: Displays the equivalent series inductance (ESL) of
the measured impedance in Henry.
Cs: Displays the equivalent series capacitance (ESC) of
the measured impedance in Farad.
Q: Displays the Q-factor of the measured impedance
(see Impedance result equations for details).
Nyquist: Displays the measured impedance in a
Nyquist chart
Q(Tg): Displays the Q-factor derived from group delay.
This value is used for NISM (see Cursor calculations for
details).
tan(δ): Displays the tan(δ) of the measured admittance.
(see Impedance result equations for details).
40
OMICRON Lab

Table of Contents