EasyManua.ls Logo

Omicron Lab Bode 100 - Page 41

Omicron Lab Bode 100
160 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Admittance Result Formats
Magnitude: Displays the magnitude of the measured
admittance in Siemens.
Magnitude (dB): Displays the magnitude of the
measured admittance in dBS.
Phase (°): Displays the phase of the measured
admittance in degrees
Phase (rad): Displays the phase of the measured
admittance in radians
Tg: Displays the group delay of the measured
admittance in seconds (see Gain result equations for
details).
Polar: Displays the measured admittance in a polar
chart.
Real: Displays the conductance of the measured
admittance in Siemens.
Imaginary: Displays the susceptance of the measured
admittance in Siemens
Rp: Displays the equivalent parallel resistance of the
measured admittance in Ohms.
Lp: Displays the equivalent parallel inductance of the
measured admittance in Henry.
Cs: Displays the equivalent parallel capacitance of the
measured admittance in Farad.
Q: Displays the Q-factor of the measured admittance
(see Impedance result equations for details).
Nyquist: Displays the measured admittance in a
Nyquist chart.
Q(Tg): Displays the Q-factor derived from group delay
(see Cursor calculations for details).
tan(δ): Displays the tan(δ) of the measured admittance.
(see Impedance result equations for details)
Measurement types and applications
OMICRON Lab
41

Table of Contents