Impedance Adapter measurement example
In this measurement example we will determine the inductance (Ls) and the equivalent series
resistance (Rs) of an inductance with the B-WIC impedance measurement adapter for through-hole-
type components. Connect the B-WIC to Bode 100 using three BNC cables as shown below.
Figure 8-12: Connecting the B-WIC to Bode 100
Now start the Bode Analyzer Suite and enter the Impedance Adapter measurement mode by clicking
Impedance Analysis and then .
Before you can start a measurement you have to perform a Full range calibration. To do so please
follow the steps described in Impedance Adapter Calibration on page 98.
After having performed the calibration please put the inductor you want to measure into the
impedance measurement adapter as shown in 8-12 on page 62.
62