The output level is calculated by the equation:
Output level = Reference level + Delta
When changing the reference level it might happen that the calculated Output level is higher than the
maximum output level of Bode 100 or lower than the minimum output level of Bode 100. In such a
case the level is automatically limited to the device limits.
This is indicated by an orange limit line and an orange cell color in the shaped level dialog as shown in
the following figure:
Figure 9-7: Output level limited to the minimum output level of the device
82