183
Error and Warning Messages
930.xxx Error [930.xxx] Measurement input Vsec defective, contact next Omicron service
center.
Reason: Power supply for analog input "Sec" is defective.
Solution: Contact your next OMICRON service center.
931.xxx Error [931.xxx] Measurement input Vprim defective, contact next Omicron service
center.
Reason: Power supply for analog input "Prim" is defective.
Solution: Contact your next OMICRON service center.
932.xxx Error [932.xxx] Data error.
Reason: The sequence of reading of the analog input channels is not correct.
Solution: If this error occurs more frequently, contact your next OMICRON service
center.
933.xxx Error [933.xxx] Hardware failure, contact next Omicron service center.
Reason: Temperature detection circuit on secondary side is defective.
Solution: Send the device to your next OMICRON service center.
Safety action: Power supply is switched off and all measurements are disabled.
934.xxx Warning [934.xxx] Reverse power, don't disconnect any wires, don't switch off mains,
wait until power is dissipated. I = xxxxA.
Reason: The device receives reverse power of > 20mA.
Solution: Wait until the power is dissipated within the device and the error message
disappears.
Safety action: All measurements are stopped until reverse power is dissipated.
935.xxx Error [935.xxx] No valid CT Analyzer software. Insert a CF-Card with valid Software and
press "Update Firmw.".
Reason: Cannot find valid CT Analyzer software.
Solution: Insert a Compact Flash card with valid software (CTAnalyzer.bin) in the
directory A:\Omicron.
936.xxx Warning [936.xxx] Corrupt Calibration Data for voltage inputs. Change to tools menu
and try to reset factory calibration. Until update of calibration data the device will be not
calibrated.
Reason: The calibration data checksum for the analog inputs is not correct.
Solution: Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
937.xxx Warning [937.xxx] Corrupt Calibration Data for power output. Change to tools menu
and try to update factory calibration. Until update of calibration data the device will be
not calibrated.
Reason: The calibration data checksum for the power output is not correct.
Solution: Try to update the factory calibration using the tools menu. If this does not solve
the problem, contact your next OMICRON service center.
938.xxx Warning [938.xxx] MIF data block 1 corrupt.
Reason: The checksum of the factory settings data is faulty.
Solution: Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.
939.xxx Warning [939.xxx] MIF data block 2 corrupt.
Reason: The checksum of the factory settings data is faulty.
Solution: Licensing or device settings data are possibly corrupt. Contact your next
OMICRON service center.