CT Analyzer User Manual
184
940.xxx Warning [940.xxx] CMOS data block corrupt. Check all device settings. Press Clear
Values to work with default values. Press OK to work with corrupt device settings.
Reason: Device settings data block corrupt.
Solution: Check all device settings.
941.xxx Error [941.xxx] Corrupt factory calibration data for voltage inputs. Contact next
Omicron service center.
Reason: Factory calibration data block for voltage inputs corrupt.
Solution: Contact your next OMICRON service center.
942.xxx Error [942.xxx] Corrupt factory calibration data for power unit. Contact next Omicron
service center.
Reason: Factory calibration data corrupt.
Solution: Contact your next OMICRON service center.
943.xxx Error [943.xxx] Error overwriting calibration data.
Reason: Factory calibration data cannot be restored.
Solution: Contact your next OMICRON service center.
944.xxx Warning [944.xxx] Corrupt user text! Change to tools menu and try to update text. Until
update of user text, the device will use default text.
Reason: Checksum for user language support data invalid.
Solution: Copy a user text file (CTUser_xxx.bin) to the directory A:\Omicron\ on the
Compact Flash card and try to update the user language support data using the
"Update Text" function in the tools menu.