EasyManuals Logo

Omicron CP TD1 Reference Manual

Omicron CP TD1
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
CP TD1 Reference Manual V 1.44
24
If you experience substantial deviations, re-calibrate the CP TD1:
Cx = Cref / Cmeas and
•DF/PF+=DFref-DFmeas
as described in 3.3 on page 29.
A re-calibration of the CP TD1 is also shown in the test report (.xml file).
Note: If you change the factory-made calibration, the responsibility for the
accuracy of the CP TD1 will be in your hands.
2.4.1 Calibration Tips
For calibration set the averaging factor to maximum and the filter bandwidth
to ± 5Hz (refer to 3.1 ”TanDelta Test Card - Main Page (1/2)” on page 25).
To reset to the factory settings, select "DF/PF+" to 0.0 ppm and "Cx" to 1.000
(refer to 3.3 ”TanDelta Test Card - Settings Page (1/2)” on page 29).

Table of Contents

Other manuals for Omicron CP TD1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CP TD1 and is the answer not in the manual?

Omicron CP TD1 Specifications

General IconGeneral
BrandOmicron
ModelCP TD1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals