CP TD1 Reference Manual V 1.44
24
If you experience substantial deviations, re-calibrate the CP TD1:
Cx = Cref / Cmeas and
•DF/PF+=DFref-DFmeas
as described in 3.3 on page 29.
A re-calibration of the CP TD1 is also shown in the test report (.xml file).
Note: If you change the factory-made calibration, the responsibility for the
accuracy of the CP TD1 will be in your hands.
2.4.1 Calibration Tips
For calibration set the averaging factor to maximum and the filter bandwidth
to ± 5Hz (refer to 3.1 ”TanDelta Test Card - Main Page (1/2)” on page 25).
To reset to the factory settings, select "DF/PF+" to 0.0 ppm and "Cx" to 1.000
(refer to 3.3 ”TanDelta Test Card - Settings Page (1/2)” on page 29).