SYMPTOMS CORRECTIVE ACTION
c) Faulty handpiece. Replace handpiece.
Diathermy System
No diathermy power output to the
handpiece when footswitch is depressed.
a) Loose connection of cable to console or to handpiece
(in case of microforceps).
Insert plug appropriately.
b) Faulty diathermy handpiece. Replace handpiece.
c) Some dirt on probe tip. Clean the probe tip
Control Console
Gas leaks from panel connector when
connecting vitrectomy handpiece.
a) Female Luer connector not properly engaged. Ensure
that connector is fully engaged by gently twist
locking it.
b) Damaged connector. Replace with new connector
Audible gas leaks from the console.
Tube fittings or valve or solenoid valve needs servicing.
Contact nearest sales/service office.
Console malfunctions.
Console malfunctions. Return console to supplier or
contact nearest sales/service office.
Phacoemulsification
Tuning not possible. U/S (Check Tip
message).
a) Loose titanium tip. Screw the tip properly into the
handpiece by using ring spanner.
b) Damaged titanium tip. Inspect the tip and replace if
necessary.
c) Damaged U/S handpiece. Replace the handpiece.
Poor U/S power. a) Insufficient U/S power preset. Increase U/S power
preset.
b) Footswitch not fully depressed. Depress the
footswitch to max travel.
c) Defective titanium tip. Replace tip.
d) Weak handpiece. Handpiece Piezoceramics normally
deteriorate with use and sterilization cycles. Send
handpiece to the OPTIKON 2000 S.P.A. Service
centre.
e) Aspiration level is too low. An insufficient aspiration
level may not hold the nucleus fragments properly
during emulsification. Increase the vacuum level
while taking care to mantain a proper fluidic
balance.