PULSAR
2
ESP
Installation and Operating Manual
OPTIKON 2000
Cod. 111007EN Rev.E
10-4
SYMPTOMS CORRECTIVE ACTION
Priming errors
Err 11: Vac offset. Some vacuum is present although aspiration pump has
not been activated.
a) Priming has been restarted after an Err 13. Open the
irrigation clamp and restart priming procedure.
b) The I/A cassette has remained connected to the
console for several hours. Replace the I/A cassette.
c) Vacuum sensor needs calibration. Contact nearest
sales/service office to recalibrate the unit.
Err 12: Major leakage There is a major leakage in the I/A system:
a) Test chamber is not installed on the sleeve. Ensure
that the test chamber is properly installed on the
sleeve.
b) I/A lines are not connected to the handpiece.
Connect the I/A lines from the cassette to the phaco
handpiece.
Err 13: No irrigation No irrigation flowing, irrigation clamp on the line from
the drip chamber is closed. Open the clamp and restart
priming procedure.
Err 14: Aspiration occluded The aspiration line is partially occluded. Handpiece and
tip may have not been properly cleaned before
sterilization. Replace tip and/or handpiece.
Err 15: Minor leakage There is a minor leakage in the I/A system:
a) I/A connectors are not fully engaged in the
handpiece connectors. Push connectors in.
b) The sleeve or the test chamber are not correctly
assembled. Verify that the test chamber is fully
engaged on the sleeve and that the sleeve is properly
positioned on the handpiece.
Weak handpiece
Handpiece efficiency is low and it cannot deliver more
than 50
µ
m. Handpiece should be serviced; contact
nearest sales/service office
VIT (Vitrectomy)
Vitrectomy handpiece will not operate
when the footswitch is depressed.
a) You are involuntarily hitting aspiration only
footswitch (left side). Press main footswitch only.
b) Handpiece connector not properly engaged. Ensure
that the connector is properly engaged.