SECTION 5 REMOTE OPERATION
84
5.4.1 OUTPUT PARAMETERS (cont)
:[SOURce:]FREQuency <n> sets Output Frequency.
n = <:SOUR:FREQ:LIM:MIN> to <:SOUR:FREQ:LIM:MAX> Hz.
Also query Frequency display with :MEASure:FREQuency?.
Returns the set value only, as UPC does not measure Frequency.
:[SOURce:]CURRent:LIMit <n> sets Output Current Limit.
n = <0> to <Current meter range> RMS Amps
:[SOURce:]PHASe2 <n> sets Output Φ angle B.
n = <0> to <359> degrees
:[SOURce:]PHASe3 <n> sets Output Φ angle C.
n = <0> to <359> degrees
:[SOURce:]WAVEFORM <n> sets Output Waveform.
n = <1> to <22>
:[SOURce:]WAVEFORM1 <n> sets Output Waveform ΦA.
n = <1> to <22>
:[SOURce:]WAVEFORM2 <n> sets Output Waveform ΦB.
n = <1> to <22>
:[SOURce:]WAVEFORM3 <n> sets Output Waveform ΦC.
n = <1> to <22>
:[SOURce:]RAMP <t> sets Output TRANSITION TIME of voltage and frequency execution.
t = <0> to <300> seconds in increments as small as .0002 Sec
:[SOURce:]FREQuency:SPAN <n> sets Output Frequency Span.
n = <150>, <300>, <600> Hz.
After sending this command, an 8 second delay is required before
sending any additional command or query. Allows UPC to restart.
This setting changes the resolution of the output signal, metering and
Harmonic Analysis option. See 2.1.1 and 4.3.1.1.
:[SOURce:]FREQuency:LIMit:MAXimum <n> sets Output Frequency max limit.
n = <15.00> to <1200> Hz., depending on Frequency Span
:[SOURce:]FREQuency:LIMit:MINimum <n> sets Output Frequency min limit.
n = <15.00> to <1200> Hz., depending on Frequency Span
:[SOURce:]VOLTage:LIMit:MAXimum <n> sets Output Voltage max limit.
n = <0.00> to <600> VAC RMS, must be greater than MIN
:[SOURce:]VOLTage:LIMit:MINimum <n> sets Output Frequency min limit.
n = <0.00> to <600> VAC RMS, must be less than MAX
:[SOURce:]VOLTage:INITial <b> sets initial output voltage at power up.
b = <0> (0 volts) or <1> (last executed)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Pacific Power Source UPC-3 and is the answer not in the manual?

Pacific Power Source UPC-3 Specifications

General IconGeneral
BrandPacific Power Source
ModelUPC-3
CategoryController
LanguageEnglish

Related product manuals