EasyManua.ls Logo

palintest Micro 600 - Mg;L or Ppm Calibration

palintest Micro 600
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3
Calibration
13
1 Perform 100% saturation calibration.
See Section 3.3.1 - 100% Calibration.
2 Rinse the probe well with DI water
or rinse solution.
3 Dip the probe into a sample of known
oxygen concentration (i.e. determined by
titration or another instrument). Wait for
the reading to stabilise.
4 Press MODE to select the mg/L (ppm).
5 Press CAL. The CAL indicator and current
concentration display briefly, before CAL
flashes. The concentration value based on
the factory default calibration is shown.
6 Press ▲▼ to adjust the reading to match
the known oxygen concentration value.
7 Press HOLD/ENTER to confirm the
calibration. The meter displays “CO”,
automatically calibrates to the entered
value, then returns to measurement mode.
NOTE: The minimum calibration value is 2
mg/L. To prevent erroneous calibrations, the
calibration is limited to ±40% adjustment
of the factory default value.
3.4 mg/L or ppm Calibration
Performing a 100% saturation calibration will simultaneously calibrate the
corresponding mg/L (ppm) concentration value. Therefore, additional mg/L
(ppm) calibration isn’t required in most circumstances.
If desired, you can perform a calibration adjustment in mg/L (ppm) without
affecting your % saturation calibration value.
Micro600DO_ZI INST 1240 16/9/15 12:57 Page 13

Related product manuals