EasyManua.ls Logo

palintest Micro 600 - Saturation Offset Adjustment (OFS.6)

palintest Micro 600
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5
Setup
25
Use this feature to offset the meter’s value
when cross referenced to another DO meter.
The Micro 600 DO allows +/- 10.0% offset
adjustment. View the offset value from the
Electrode Data menu “ELE.3”.
1 Using the % saturation mode of your
Micro 600 DO, observe the reading a
sample solution after it has stabilised.
2 Similarly, observe the reading of the
same sample using another DO meter as
a reference. The probe of the reference
meter should be immersed in the same
sample at the same depth.
3 Switch off the Micro 600 DO and enter
the % saturation setup menu.
4 Press ▲▼ to select “OFS.6”.
5 Press HOLD/ENTER to enter the menu. The
display will momentarily show the measured
reading based on the last calibration before
displaying the last offset adjusted value.
6 Use ▲▼ to enter the value of the
reference DO meter.
7 Press HOLD/ENTER to confirm.
8 Press CAL to return to measurement mode.
Note: User calibrations will reset the
offset adjustment to 0.0%.
5.8 % Saturation Offset Adjustment (OFS.6)
6 If “NO”, the User Reset menu “b.USR” is displayed.
7 Press HOLD/ENTER to enter the menu.
8 Use ▲▼ to select YES (reset user) or NO (do not reset user).
9 Press HOLD/ENTER to confirm.
10 If “YES”, the meter returns to measurement mode after resetting.
11 Press CAL to return to measurement mode.
Micro600DO_ZI INST 1240 16/9/15 12:57 Page 25

Related product manuals