9
Factory Default Settings
32
Configuration Setup Menu
COF.1
YES / NO YESAutomatic Temperature Compensation
A.ATC
mg/L or ppm mg/LDO Concentration Units
b.DO
View only _ _ _ _Calibration Data
CAL.2
Electrode Data Menu
ELE.3
View only 1.000Electrode Slope Factor
FACT
View only 0.0%Viewing the % Saturation offset adjustment
OFS
View only 50 mVView mV Value at 100% Saturation
HI.mV
View only 0 mVView mV Value at 0% Saturation
LO.mV
YES / NO YESAutomatic Power Off
AtO.4
Reset to Factory Default Menu
rSt.5
NO / YES NOSelection of Calibration Reset
A.CAL
NO / YES NOSelection of User Reset
b.USR
+/- 10.0% 0.0%% Saturation Offset Adjustment
OFS.6
Dissolved Oxygen Parameters Menu
DPr.7
500 - 1499 760Pressure Adjustment in mm Hg
A.HG
66.6 - 199.9 101.3Pressure Adjustment in kPa
A.PA
0.0 - 50.0 0.0Salinity Adjustment in ppt
b.SAL
0.0 - 50.0 °C 25.0°CManual Temperature Compensation
Menu Description Options Default
Micro600DO_ZI INST 1240 16/9/15 12:57 Page 32