For part numbers, see Section 11 - REPLACEMENTS AND ACCESSORIES.
1 Unscrew the cap counter clockwise from the probe sensing tip.
2 Rinse the probe under running water.
3 Mount the nozzle tip onto the syringe provided. Fill the syringe with the
refill solution through the tip of the plastic bottle.
4 Hold the probe upside down. Insert the nozzle tip into one of the 4 holes
surrounding the silver cathode. Inject the fill solution into the probe body
until solution leaks out from the fill hole (approximately 5mL).
5 Replace pre-assembled cap by tightening clockwise until hand tightened.
6 Allow at least 1 hour for the electrode to equilibrate before usage.
6
Probe Maintenance
29
6.2 Electrolyte Solution
The electrolyte solution in your probe’s cap will deplete on usage and will
need to be replaced periodically. The replacement electrolyte solution included
with your probe comes premixed and ready to use. To order more electrolyte
solution, Section 11 - REPLACEMENTS AND ACCESSORIES.